Reference · Quality

Purity, content and what a 10 mg label actually means

Three numbers describe what is in a vial. Most certificates show one of them, some show two, and almost none show all three. Here is what each one measures and what it does to the concentration you calculate.

Published 15 September 2026 · Last reviewed 15 September 2026

Lyophilised research peptide vials and a laboratory balance, Global Bioanalytical Laboratories

Two numbers sit beside each other on most test reports and they look like they are measuring the same thing. They are not. One tells you how clean the material is, the other tells you how much of it is there, and reading one as though it were the other is the most common mistake made with a certificate of analysis. There is also a third number, the one that actually answers the question people think they are asking, and it is missing from nearly every document in this industry. This article covers all three, using reports published on our own Certificates of Analysis page rather than typical figures, because a real report is more useful than a tidy one.

Two numbers that look like one

A purity figure is a ratio. A content figure is a mass. They can move independently of each other, and a vial can be exceptionally clean and still hold less peptide than its label suggests, or hold more. Purity says nothing about quantity and quantity says nothing about cleanliness.

The label itself is a third thing again. A figure such as 10 mg on a vial is a nominal fill, which means the quantity the production line was targeting. It is a specification, not a measurement of that vial. The only way to know what a given vial holds is for a laboratory to quantify it, and the results below show the label being missed in both directions.

In plain English

Purity is a percentage, so it answers “how clean”. Content is a weight, so it answers “how much”. The number printed on the label answers neither; it is what the vial was meant to be filled with.

What HPLC purity measures

High performance liquid chromatography, usually shortened to HPLC, pushes a dissolved sample through a column packed with a stationary material and measures what comes out the other end over time. Components separate because they interact with that packing to different degrees, and a detector at the end records absorbance as each one passes. For peptides the detector is normally set at 214 nm, where the peptide bond itself absorbs, so every peptide species in the sample is seen whether or not it contains an aromatic residue.

The purity figure is then the area of the target peak expressed as a percentage of the total detected area. A result of 99.076% means that of everything the instrument saw, just over 99 parts in 100 was the target peptide and the remainder was related material: truncated sequences that stopped short during synthesis, deletion sequences missing a residue, oxidised or deamidated forms, and synthesis by products.

The important limitation is in the word “detected”. Anything the method does not see is not in the denominator. Water, counterion, residual solvent and inorganic salts absorb little or nothing at 214 nm, so they are largely invisible to the measurement. A purity percentage is a statement about the peptide material in the vial, not about the contents of the vial.

In plain English

Purity compares peptide against other peptide. It is a fair fight between the molecule you wanted and the near misses from the same synthesis. The water and salt sitting in the powder are not in the contest at all.

What content per vial measures

Content is the quantity of the named peptide found in the vial, reported in milligrams. It is measured by running the sample against a reference standard of known concentration, so the instrument is comparing peak area to a calibration rather than to the other peaks in the same run. That is why a report can state both a purity percentage and a milligram figure: they come from different comparisons.

Content is the number that matters for concentration work, because it is the only one expressed in mass. A purity percentage cannot be converted into milligrams, no matter how high it is.

Net peptide content, the missing number

Net peptide content is peptide expressed as a proportion of the total mass of powder in the vial. It is the number that explains why a vial weighing 12 mg can hold 10 mg of peptide, and it is absent from most certificates in this industry.

Lyophilised peptide is not pure peptide by mass. After freeze drying, a vial typically holds the peptide itself, a counterion paired to the basic residues, water both bound to the peptide and left over from the process, and residual solvent or inorganic salt from purification. The proportions vary with the sequence and the purification route, and the ranges below are the ones ordinarily encountered rather than figures from any particular vial.

ComponentTypical share of vial massWhere it comes from
Peptide70 to 90%The synthesis itself
Counterion5 to 20%Trifluoroacetate from purification, or acetate after exchange
Water2 to 10%Bound to the peptide and residual from lyophilisation
Salts and residual solvent0 to 5%Buffers and mobile phase carried through

Ordinary ranges for synthetic peptides supplied as a lyophilised powder. Any individual vial is described only by its own analysis.

In plain English

If you tipped a vial onto a balance, a meaningful slice of what you weighed would not be peptide. Net peptide content is the size of that slice, stated the other way round. Purity will not tell you it is there, because purity never weighed the vial.

Why counterion load changes with the sequence

Most synthetic peptides are purified by reversed phase chromatography using trifluoroacetic acid in the mobile phase, and the peptide comes off that process as a trifluoroacetate salt. Trifluoroacetate pairs with positively charged sites, which in a peptide means the basic residues: arginine, lysine and histidine, plus the free amino terminus.

The consequence is that counterion load is a property of the sequence rather than a constant. A peptide carrying several arginine and lysine residues binds proportionally more trifluoroacetate than a neutral sequence of similar length, so two vials of identical label weight and identical purity can hold materially different quantities of peptide. Where trifluoroacetate is unsuitable it can be exchanged for acetate, which changes the mass balance again because the two counterions differ in molecular weight.

Three vials, three answers

The reports below are published in full, with verification keys, on our Certificates of Analysis page. They were produced by Janoshik Analytical on vials from the production batches we supply, and each one can be checked at the laboratory’s own verification page without taking our word for anything.

ReportLabelMeasured contentPurityAgainst label
Retatrutide #20030310 mg11.34 mg99.076%13% above
MOTS-c #10476310 mg9.85 mg99.748%1.5% below
Tesamorelin #17260610 mg10.05 mg98.590%0.5% above

Analytical results relate only to the samples tested. Batch to batch variation is possible.

Three vials all labelled 10 mg, holding 11.34 mg, 9.85 mg and 10.05 mg. Note also that the highest purity of the three sits on the vial holding the least peptide, which is the clearest demonstration available that the two numbers are independent.

The GHK-Cu report, #225308, shows the mass balance problem more directly than any of them, because copper peptides force the laboratory to separate the components. A vial labelled 100 mg returned 96.94 mg of GHK-Cu, and the laboratory then split that figure into 81.13 mg of GHK peptide and 15.81 mg of copper. Both numbers are correct and they answer different questions. If the assay is designed around the complex, 96.94 mg is the relevant mass. If it is designed around the peptide, the figure is 81.13 mg, which is roughly 81% of the label.

Carried into concentration, a vial reconstituted to a stock solution of 2 mL gives 5.67 mg/mL from the measured retatrutide content against the 5.00 mg/mL the label implies. That gap is 13%, and it appears in every downstream dilution. The molarity calculation and a molecular weight table for the catalogue are covered in the article on molarity and concentration, and the arithmetic is built into Peptide Pro.

In plain English

Four real vials, four different relationships between the label and the contents. Two over, one under, and one where most of what you bought is peptide and the rest is copper that belongs there. None of this indicates a problem with any of the vials. It indicates that a label is a target and a measurement is a measurement.

How each number is determined

Purity and content both come from chromatography, purity as an area ratio and content against a calibrated reference standard. The other components need separate methods. Water is determined by Karl Fischer titration, a technique that reacts specifically with water and is unaffected by the rest of the powder. Peptide content by mass is established by amino acid analysis, in which the peptide is hydrolysed back to its constituent amino acids and those are quantified, or by ultraviolet absorbance at 280 nm where the sequence contains tryptophan or tyrosine. Counterion is measured by ion chromatography or by fluorine nuclear magnetic resonance in the case of trifluoroacetate.

The reason net peptide content rarely appears is cost. Each of those determinations is a separate submission on a separate instrument, and a purity and identity run is what most of the supply chain pays for. The article on HPLC and mass spectrometry covers the instrumentation in detail.

What to ask for when the number is absent

If a certificate shows purity alone, it has told you how clean the material is and nothing about how much of it there is. Content in milligrams is the next thing to ask for, and it is the figure that makes a concentration calculation defensible. Net peptide content is worth requesting where the work is quantitative and the sequence is strongly basic, since that is where the gap between vial mass and peptide mass is widest.

Certificates of analysis, where available, are published on our Certificates of Analysis page, and where an independent report exists it is published there in full with its verification key. Reading either document is covered in the guide to reading a certificate of analysis.

Common questions

Does a 10 mg vial contain 10 mg of peptide?

Not necessarily, and not reliably in either direction. The label states a nominal fill target. Of three vials labelled 10 mg tested for us, one held 11.34 mg, one held 9.85 mg and one held 10.05 mg. Only a quantification of that batch answers the question.

What is the difference between purity and net peptide content?

Purity is the proportion of detected peptide material that is the target sequence, expressed as a percentage. Net peptide content is the proportion of the total powder mass that is peptide at all, with the remainder being counterion, water and residual salts. A sample can be 99% pure and still be 80% peptide by mass.

Why is net peptide content missing from most certificates?

Because it requires separate analyses on separate instruments, including Karl Fischer titration for water and amino acid analysis or ultraviolet quantification for peptide mass. Most certificates report the purity and identity work that the supply chain routinely commissions.

Does a higher purity percentage mean more peptide in the vial?

No. The two are independent. Among the reports published on our Certificates of Analysis page, the highest purity figure belongs to the vial holding the least peptide against its label.

Which figure should be used to calculate a stock concentration?

Measured content in milligrams, where a report provides it, since concentration is a mass divided by a volume. Where only the label figure is available, the calculation carries whatever error sits between the nominal fill and the actual contents.

None of this is a reason to distrust a label. It is a reason to know what a label is. A nominal fill figure is the production line stating its target, and a test report is a laboratory stating what it found, and the useful habit is simply keeping the two apart in your head.

About this article

Written by the research team at Global Bioanalytical Laboratories, Adelaide, South Australia. We revise these pages as new analytical data comes in, and the Last reviewed date above records when this one was last checked. If something here is wrong or has moved on, tell us at lab-tech@globalbioanalytical.com.au and we will correct it.

References

  1. Global Bioanalytical Laboratories, Certificates of Analysis: reports #200303, #172606, #225308 and #104763, with verification keys. https://globalbioanalytical.com.au/quality-assurance/
  2. Janoshik Analytical, report verification service. https://janoshik.com/verify/
  3. Andrushchenko VV, Vogel HJ, Prenner EJ. Optimization of the hydrochloric acid concentration used for trifluoroacetate removal from synthetic peptides. Journal of Peptide Science, 2007;13(1):37-43. https://doi.org/10.1002/psc.793
  4. Roux S, Zekri E, Rousseau B, Paternostre M, Cintrat JC, Fay N. Elimination and exchange of trifluoroacetate counter-ion from cationic peptides: a critical evaluation of different approaches. Journal of Peptide Science, 2008;14(3):354-359. https://doi.org/10.1002/psc.951

Analytical results relate to the samples tested and batch to batch variation is possible. Test reports describe the analytical characteristics of research materials only.