Research reference

HPLC and mass spectrometry: how a research peptide is verified, and what each instrument cannot tell you

Two instruments answer two different questions. High performance liquid chromatography reports how clean a peptide is; mass spectrometry reports what it is. Neither says how much of it there is on its own, and neither confirms the order of the residues. What happens inside each instrument, how to read the number it produces, where the published reports on our own products fit, and what the open market looked like when someone finally ran the instruments on it.

Published 13 October 2026 · Last reviewed 13 October 2026

High performance liquid chromatography instrument and sample vials in a laboratory, Global Bioanalytical Laboratories
Read this first

A certificate of analysis for a research peptide rests on two instruments. High performance liquid chromatography, HPLC, pushes the dissolved sample through a packed column so that its components come out one at a time, and a detector records each one as a peak; the purity figure is the share of the total that belongs to the target peptide. Mass spectrometry turns the peptide into charged ions and weighs them, and a measured mass that matches the expected one is the evidence that the molecule is the right one. Purity without identity tells you how clean something is without saying what it is. Identity without purity tells you the right molecule is present without saying how much of what came with it. Neither instrument, run alone, gives the quantity of peptide in the vial, which needs a reference standard, and neither confirms the order of the amino acids, which needs a further experiment again.

This reference explains what each instrument does, how its number is produced and read, and where it stops, using the independent reports published on our Certificates of Analysis page as the worked examples. Nothing here is a protocol. Everything describes analytical work on research materials in a laboratory.

Why a certificate needs both instruments

There is a habit in this market of reading a purity percentage as if it settled everything, and it settles exactly one thing. A vial can be 99 per cent pure and contain the wrong peptide, because purity is a ratio of the main component to everything else the detector saw, and the detector does not know what the main component is. The clearest demonstration we can offer is on our own Certificates of Analysis page, where the two retatrutide reports were run as a blind identity test: the laboratory was asked to determine which of semaglutide, tirzepatide and retatrutide the vials contained, and confirmed retatrutide in both sizes, with purities of 99.076 per cent for the 10 mg vial and 99.689 per cent for the 30 mg vial.1 Those three peptides belong to one class, look alike on a chromatogram, and weigh 4,113.6, 4,813.5 and 4,731.4 g/mol respectively, so the mass spectrometer tells them apart at a glance while the purity figure, on its own, could not have told anyone which of the three was in the vial. The guide to reading a certificate of analysis covers what each field on the document proves; this reference is about the instruments behind the two fields that matter most.

In plain English

Purity is how clean. Identity is what. A 99 per cent figure on the wrong molecule is still the wrong molecule, which is why a certificate that shows a purity and never confirms a mass has only done half the job. Our own retatrutide reports were run blind against two near relatives for exactly this reason, and the mass is what gave the answer.

What happens on the column

Peptide purity is almost always measured by reversed phase HPLC. The column is a steel tube packed with silica particles a few micrometres across whose surfaces are bonded with eighteen carbon hydrocarbon chains, the C18 phase, so the packing is hydrophobic; for peptides the particles are made with wide pores, about 300 ångströms, so that a chain of several thousand daltons can reach the surface inside them rather than only the outside.2 The sample is dissolved and loaded into a stream of mobile phase that starts as mostly water and becomes progressively richer in acetonitrile over the run, which is the gradient. Each peptide binds the hydrophobic surface when the mobile phase is watery and lets go when the acetonitrile reaches the fraction that competes for it, so the components leave the column in order of their hydrophobicity: a small polar tripeptide such as GHK elutes early and can be hard to separate from the salts that leave with the solvent front, while a long peptide carrying a fatty acid side chain, such as the GLP class, holds on until late in the gradient.2

The mobile phase also contains about 0.1 per cent trifluoroacetic acid. It does two jobs. It holds the pH near 2, so that acidic side chains are protonated and the peptide behaves consistently, and it pairs with the positive charges on lysine, arginine and the N terminus, which masks them and makes the peptide bind more predictably and leave as a sharper peak.2 That same trifluoroacetate is why a lyophilised peptide usually arrives as a trifluoroacetate salt, and why the purity and content reference has to explain that the counterion is part of the powder but never part of the purity figure. Retention time, the number of minutes a peptide takes to appear, is a property of the peptide, the column, the gradient and the temperature together, so it is a useful fingerprint only when a certificate also states the method it was produced with.

In plain English

The column is a long tube of greasy beads. Peptides stick to the grease and are washed off one at a time as the solvent gets less watery, the least greasy first and the greasiest last. The acid in the solvent keeps everything behaving the same way from run to run. How long a peptide takes to come off means nothing unless you also know the column and the solvent programme it came off.

What the detector sees

As each component leaves the column it passes through an ultraviolet detector, and the wavelength chosen decides what is visible. The peptide bond itself absorbs strongly near 214 nm, so at that wavelength every peptide produces a signal roughly in proportion to how many residues it has, and Kuipers and Gruppen tabulated the contribution of each amino acid at 214 nm precisely so that peak areas could be related to quantity.3 At 280 nm the picture is different: absorbance there comes almost entirely from the side chains of tryptophan and tyrosine, with a small contribution from cystine, which is the basis on which protein concentration is routinely measured and which Pace and colleagues quantified residue by residue.4 A peptide with neither tryptophan nor tyrosine is nearly invisible at 280 nm, and that describes a surprising share of this shelf: BPC-157, KPV, GHK, epithalon, selank, SNAP-8 and thymosin beta 4 contain no tryptophan and no tyrosine, so a purity run on any of them has to be read at 214 or 220 nm, whereas MOTS-c, kisspeptin-10, Mod GRF 1-29 and the three GLP class peptides carry aromatic residues and show at both wavelengths.

The wavelength also defines what is left out. Water, inorganic salts, trifluoroacetate and residual solvents absorb little or nothing near 214 nm, so they contribute no peak and no area, and a purity percentage is therefore a statement about the peptide material in the sample and not about the contents of the vial.3 That is the whole reason a 99 per cent pure vial can be 80 per cent peptide by mass, and it is covered field by field in the purity and content reference.

In plain English

The detector is a lamp. At 214 nm it sees every peptide bond, so every peptide shows up. At 280 nm it sees only two particular amino acids, so a peptide without them is a ghost. And it never sees salt or water at all, which is why purity and the weight of the powder are two different conversations.

Reading a purity figure

The purity figure is the area of the main peak divided by the sum of all peak areas on the chromatogram, expressed as a percentage. The reports on our Certificates page show what the output looks like in practice: 99.076 and 99.689 per cent for the two retatrutide vials, 99.748 per cent for MOTS-c, 99.341 per cent for the GHK tripeptide in the GHK-Cu report, and 98.590 per cent for tesamorelin batch FSD0716.1 Each of those is a ratio of peptide to peptide, with everything the detector could see in the denominator and nothing it could not. Three decimal places are the instrument’s integration speaking rather than a claim of that precision; the figure moves with the integration settings and the baseline, and a certificate that shows the chromatogram lets a reader see how the baseline was drawn and whether the main peak is symmetrical.

What sits in the other one or two per cent is a family of related molecules, and D’Hondt and colleagues catalogued them for peptide medicines in a review that is the reference for the field.5 Deletion sequences are missing one residue because a coupling step failed during synthesis, and they elute close to the parent; for BPC-157, a missing proline is a peak 97.1 daltons lighter. Truncated sequences stopped short at one end. Oxidation adds 16 daltons, almost always on methionine, which MOTS-c, SNAP-8 and thymosin beta 4 all carry, and which the storage and stability reference describes as the first reaction to watch for in storage. Deamidation of asparagine or glutamine adds just under one dalton and produces a peak that elutes close to the parent, and the isomerisation of aspartate produces a species of identical mass that only the chromatogram can separate. Trifluoroacetate adducts add 114. A shoulder on the main peak is the usual sign that one of these is being counted as product, and it is the reason the chromatogram is worth more than the number printed beside it.5 The certificate guide reproduces the grade table that Bachem publishes, which maps purity to use: above 95 per cent for quantitative reference work, 90 to 95 per cent for quantitative enzyme and receptor assays, and lower grades for qualitative work only.6

One assumption is built into every area percentage. It treats equal peak areas as equal amounts of material whichever peak they belong to, which is close to true for related peptides at 214 nm, because they share nearly the same residues, and less true for an impurity with a very different composition. For a research certificate that assumption is accepted and rarely stated; a validated pharmaceutical method would quantify named impurities against their own standards, which is a different and more expensive exercise.5, 11

In plain English

Purity is the big peak as a share of all the peaks. The small peaks are the near misses from the synthesis, each a predictable number of daltons heavier or lighter than the real thing, and a bump on the side of the big peak means one of them is hiding inside it. Ask for the picture, not just the percentage.

From purity to content

A purity run cannot produce a mass in milligrams, because it compares peaks within one sample and never against anything of known quantity. Content comes from a second use of the same instrument: a reference standard of the peptide at known concentration is run to calibrate the detector response, and the sample’s main peak area is read against that calibration to give the amount of peptide in the vial.2, 3 The retatrutide report #200303 is the example: a vial labelled 10 mg returned a measured content of 11.34 mg beside its 99.076 per cent purity, two numbers from two comparisons, one against the other peaks and one against a standard.1 Where no reference standard exists for a compound, which is the case for much of what the research market sells, a laboratory may calibrate against a related peptide or against the predicted absorbance from the residue values at 214 nm, and a certificate should say which.3 The molarity and concentration reference takes the measured content from there to a concentration.

Mass spectrometry: how a peptide becomes a number

A mass spectrometer measures the mass to charge ratio of ions in a vacuum, so the first job is to turn a peptide in solution into ions in the gas phase without breaking it. Electrospray ionisation, for which John Fenn shared the 2002 Nobel Prize in Chemistry, does this by spraying the solution from a fine needle at high voltage; the droplets shrink as the solvent evaporates until the peptide molecules are released carrying one or more protons picked up from the acidic solvent.7 The instrument then reports each ion at m/z, the mass of the molecule plus the mass of its protons, divided by the number of protons. For a small peptide one proton is typical, and the number reads almost directly as the molecular mass plus one: GHK appears near m/z 341.2, KPV near 343.2 as the free acid or 342.3 as the C terminal amide, and that one dalton difference is how a certificate settles which form of KPV was made.8

A large peptide picks up several protons, so it appears not as one peak but as a ladder of peaks, one for each charge state. Semaglutide, at 4,113.6 g/mol, shows ions near m/z 1,372, 1,029, 824 and 687 for the three, four, five and six proton forms. Any two adjacent rungs of the ladder are enough to solve for the charge and the mass, and the method for doing so, published by Mann, Meng and Fenn in 1989, is what every instrument’s deconvolution software still does: it collapses the ladder into a single molecular mass.9 That is the figure a certificate reports as the observed mass, and the TB-500 reference is the clearest case of why it matters, because full length thymosin beta 4 deconvolutes to about 4,963 while the seven residue fragment sold under the same name appears as a singly charged ion near 889.5, with no overlap between the two.

The other common ionisation method, matrix assisted laser desorption ionisation, MALDI, mixes the peptide into a crystalline organic matrix and fires a laser at it; the matrix absorbs the energy and lifts the peptide into the gas phase, mostly as singly charged ions, so the spectrum reads the mass almost directly without a ladder to deconvolute.10 It is fast and tolerant of salts, which is why it is widely used for checking synthetic peptides, and less suited to measuring purity.

One more distinction decides whether an observed mass matches. The average molecular weight used for concentration arithmetic is a weighted average over the natural isotopes of every atom; the monoisotopic mass counts only the lightest isotope of each. For KPV they differ by 0.2 daltons, for BPC-157 by 0.9, and for retatrutide by about 3, from 4,731.4 to 4,728.5. An instrument that resolves the individual isotope peaks reports the monoisotopic value and a certificate should say so; one that does not reports an average. A three dalton disagreement between a certificate and a database is usually this, and the molarity reference explains the two figures side by side.

In plain English

The mass spectrometer sprays the peptide into a vacuum with a few extra protons attached and weighs the result. Small peptides come through as one number; big ones come through as a ladder of numbers that the software folds back into one. The answer is the molecular mass, and if it matches what the sequence predicts, the molecule is what it says it is. Check whether the certificate is quoting the everyday average mass or the lightest isotope version before you call a three dalton gap a problem.

What a mass does not prove

A matching mass is strong evidence of the intended composition and no evidence of the order. Every arrangement of the same residues weighs the same, so a scrambled sequence, or a sequence with two residues swapped, is invisible to a molecular mass. Confirming order takes tandem mass spectrometry, in which ions of the peptide are selected, broken along the backbone, and the fragments weighed; the ladder of fragment masses reads the sequence from each end, and Steen and Mann’s account of how those fragment series are interpreted remains the standard introduction.12 Routine research certificates report a molecular mass and not a fragmentation spectrum, which is reasonable for a synthesis that followed a known sequence and worth knowing all the same.

Stereochemistry is invisible too. A D amino acid weighs exactly what its L counterpart weighs, so Mod GRF 1-29, whose second residue is D alanine, would give the same observed mass if it had been made with L alanine at that position, and ipamorelin, which carries two D residues, is the same case. The Mod GRF 1-29 and ipamorelin reference explains why those residues are there; confirming them takes chiral amino acid analysis, not a mass. Amino acid analysis itself, in which the peptide is hydrolysed and its residues counted, returns composition and not sequence, as the certificate guide says plainly. And none of these instruments reports what came with the peptide: the counterion, the water and the endotoxin are separate measurements on separate instruments, and the certificate guide covers each of them.

In plain English

A mass proves the right atoms are present in the right number. It does not prove they are in the right order, and it cannot tell a left handed amino acid from a right handed one. For a synthesis that followed the sequence those are small risks, but they are real, and a certificate that reports a mass has not proved either.

Running the two together

Chromatography and mass spectrometry are often coupled, with the column’s outflow sprayed straight into the mass spectrometer so that every peak is identified as it elutes, which is how a laboratory assigns a shoulder to a deletion sequence rather than guessing. The coupling has a cost. Trifluoroacetic acid, the acid that makes reversed phase separations of peptides sharp, suppresses the electrospray signal severely, by pairing with the peptide’s charges in the droplet; Apffel and colleagues measured the loss in 1995 and showed that formic acid, which is a weaker ion pairing agent, gives a far stronger signal at the price of broader peaks.13 The practical consequence is that a purity method and an identity method are often two methods, one with trifluoroacetic acid read by the ultraviolet detector and one with formic acid read by the mass spectrometer, and a certificate that quotes both a purity and an observed mass has usually run both. Two compounds eluting at the same moment also suppress each other’s signal in the spray, which the peptide compatibility reference describes with a published example, and which is one more reason a method on a blend such as the ones in the GLOW and KLOW reference has to be shown to separate its components before its numbers mean anything.

What a certificate should say about its methods

For most of the compounds on this shelf there is no pharmacopoeial monograph, which means no external specification exists that a batch can be measured against, and the certificate is therefore the specification as well as the result. That places weight on the method details, and the ones worth looking for are these. For the purity figure: the column type and dimensions, the gradient, the flow rate, the detection wavelength, and ideally the chromatogram itself. For the identity: the ionisation method, the observed mass, the expected mass it was compared with, and whether the two are average or monoisotopic values. For content: the reference standard used, or the calibration approach if there was none. For every field: the batch number that matches the vial, the date and the laboratory. In regulated manufacturing these methods are validated against the ICH Q2(R2) guideline, which sets out how specificity, accuracy, precision, linearity and range are demonstrated before a method is relied on, and a contract laboratory working to that standard will say so.11 Research certificates rarely carry a validation statement, and that is not a defect so much as a fact about the category; a report that can be verified against the laboratory’s own records, as the Janoshik reports on our Certificates page can be through their verification keys, is the nearest thing the research market has to traceability.1

In plain English

With no official standard to measure against, the certificate is the standard, so it has to show its working: which column, which solvent programme, which wavelength, which instrument, which mass it expected and which it found, and what the content was measured against. A number with none of that around it is a number you cannot check.

What the instruments found on the open market

The point of all this stops being abstract the moment the instruments are pointed at products bought without documentation. In 2024 Ashraf and colleagues test purchased semaglutide from online sellers operating without a prescription and ran the full battery: packaging failed between 59 and 63 per cent of the visual inspection criteria, measured semaglutide purity ranged from 7.7 to 14.37 per cent against the 99 per cent claimed on the labels, the measured amount of semaglutide exceeded the label by about 29 to 39 per cent, and endotoxin was present in every sample.14 Read against the sections above, that is HPLC reporting a purity an order of magnitude below the claim, calibrated content reporting more drug than labelled, and a third instrument reporting what neither of the first two could see. A decade earlier, the Ghent doping control laboratory analysed a product sold as TB-500, found by mass spectrometry that it contained the acetylated seven residue fragment of thymosin beta 4 rather than the protein, and synthesised the fragment to prove it.15 Neither finding required a new technique. Both required someone to run the ordinary ones and publish the result, and both are the reason a certificate that shows its chromatogram and its observed mass is worth reading closely before any experiment is designed around the vial.

Common questions

What does HPLC purity actually measure for a peptide?

The area of the main peak as a percentage of the total peak area on a reversed phase chromatogram, usually read by an ultraviolet detector near 214 nm where the peptide bond absorbs. It compares the target peptide with the related peptide species that came out of the same synthesis, such as deletion and truncated sequences and oxidised or deamidated forms. Water, salts and the trifluoroacetate counterion do not absorb at that wavelength, so they are not part of the figure.

Why does a certificate need mass spectrometry as well as HPLC?

Because purity is a ratio and says nothing about what the main component is. Mass spectrometry weighs the molecule and compares the observed mass with the mass the intended sequence predicts, which is what identifies it. The retatrutide reports on our Certificates of Analysis page were run as a blind test against semaglutide and tirzepatide, which weigh 4,113.6 and 4,813.5 g/mol against retatrutide’s 4,731.4, and the mass is what told them apart.

What is a charge state envelope?

A large peptide picks up several protons in electrospray ionisation, so it appears in the spectrum as a ladder of peaks, one for each number of protons, each at the molecular mass plus the protons divided by the number of protons. Semaglutide appears near m/z 1,372, 1,029, 824 and 687 for its three, four, five and six proton ions. Software deconvolutes the ladder into a single molecular mass, which is the figure reported as the observed mass.

Why does the certificate’s mass differ from the molecular weight by a few daltons?

Usually because one figure is the average molecular weight, a weighted average over the natural isotopes of every atom, and the other is the monoisotopic mass, which counts only the lightest isotope of each. For retatrutide the two are 4,731.4 and about 4,728.5. An instrument that resolves the isotope peaks reports the monoisotopic value; one that does not reports an average. The certificate should state which it is quoting.

Does a matching mass prove the sequence is correct?

No. Every arrangement of the same residues has the same mass, and a D amino acid weighs the same as its L form, so a molecular mass confirms composition rather than order or stereochemistry. Confirming order takes tandem mass spectrometry, in which the peptide is fragmented along its backbone and the fragment masses read as a sequence, and confirming stereochemistry takes chiral amino acid analysis. Routine research certificates report a molecular mass only.

Why is purity run with trifluoroacetic acid and identity with formic acid?

Trifluoroacetic acid gives sharp, well separated peptide peaks on a reversed phase column but suppresses the electrospray signal in a mass spectrometer by pairing with the peptide’s charges. Formic acid suppresses the signal far less but gives broader peaks. Laboratories therefore often run a purity method read by the ultraviolet detector with trifluoroacetic acid and a separate identity method read by the mass spectrometer with formic acid, which is why a certificate that reports both has usually run two methods.

Tools and documents

The independent reports used as examples on this page are published on the Certificates of Analysis page with their verification keys. The guide to reading a certificate of analysis explains what each field proves, the purity and content reference separates purity from the amount in the vial, and the Peptide Pro platform carries the molecular specifications and concentration arithmetic for the compounds we supply.

About this reference

Written and reviewed by the research team at Global Bioanalytical Laboratories, Adelaide, South Australia. Every figure in this article is taken from the published source cited beside it or computed from the molecular formulas in the molarity reference. Corrections and updates: lab-tech@globalbioanalytical.com.au.

References

  1. Global Bioanalytical Laboratories. Certificates of Analysis: independent reports #200303 and #200305 (retatrutide, blind identity and purity), #104763 (MOTS-c), #172606 (tesamorelin, batch FSD0716) and #225308 (GHK-Cu, batch 2026.08.10), with verification keys. globalbioanalytical.com.au/quality-assurance; Janoshik Analytical, report verification: janoshik.com/verify
  2. Mant CT, Chen Y, Yan Z, Popa TV, Kovacs JM, Mills JB, Tripet BP, Hodges RS. HPLC analysis and purification of peptides. Methods in Molecular Biology. 2007;386:3–55. doi:10.1007/978-1-59745-430-8_1
  3. Kuipers BJH, Gruppen H. Prediction of molar extinction coefficients of proteins and peptides using UV absorption of the constituent amino acids at 214 nm to enable quantitative reverse phase high-performance liquid chromatography–mass spectrometry analysis. Journal of Agricultural and Food Chemistry. 2007;55(14):5445–5451. doi:10.1021/jf070337l
  4. Pace CN, Vajdos F, Fee L, Grimsley G, Gray T. How to measure and predict the molar absorption coefficient of a protein. Protein Science. 1995;4(11):2411–2423. doi:10.1002/pro.5560041120
  5. D’Hondt M, Bracke N, Taevernier L, Gevaert B, Verbeke F, Wynendaele E, De Spiegeleer B. Related impurities in peptide medicines. Journal of Pharmaceutical and Biomedical Analysis. 2014;101:2–30. doi:10.1016/j.jpba.2014.06.012
  6. Bachem. Quality control of amino acids and peptides: a guide. Bachem Knowledge Center. bachem.com
  7. Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM. Electrospray ionization for mass spectrometry of large biomolecules. Science. 1989;246(4926):64–71. doi:10.1126/science.2675315
  8. PubChem, US National Library of Medicine. Lys-Pro-Val, compound CID 125672: molecular weight 342.43 g/mol, monoisotopic mass 342.23 Da. pubchem.ncbi.nlm.nih.gov
  9. Mann M, Meng CK, Fenn JB. Interpreting mass spectra of multiply charged ions. Analytical Chemistry. 1989;61(15):1702–1708. doi:10.1021/ac00190a023
  10. Karas M, Hillenkamp F. Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltons. Analytical Chemistry. 1988;60(20):2299–2301. doi:10.1021/ac00171a028
  11. International Council for Harmonisation. ICH Q2(R2) Validation of analytical procedures, Step 5, effective 14 June 2024. European Medicines Agency scientific guideline page. ema.europa.eu
  12. Steen H, Mann M. The ABC’s (and XYZ’s) of peptide sequencing. Nature Reviews Molecular Cell Biology. 2004;5(9):699–711. doi:10.1038/nrm1468
  13. Apffel A, Fischer S, Goldberg G, Goodley PC, Kuhlmann FE. Enhanced sensitivity for peptide mapping with electrospray liquid chromatography-mass spectrometry in the presence of signal suppression due to trifluoroacetic acid-containing mobile phases. Journal of Chromatography A. 1995;712(1):177–190. doi:10.1016/0021-9673(95)00175-M
  14. Ashraf AR, Mackey TK, Vida RG, Kulcsár G, Schmidt J, Balázs O, Domián BM, Li J, Csákó I, Fittler A. Multifactor quality and safety analysis of semaglutide products sold by online sellers without a prescription. Journal of Medical Internet Research. 2024;26:e65440. doi:10.2196/65440
  15. Esposito S, Deventer K, Goeman J, Van der Eycken J, Van Eenoo P. Synthesis and characterization of the N-terminal acetylated 17-23 fragment of thymosin beta 4 identified in TB-500, a product suspected to possess doping potential. Drug Testing and Analysis. 2012;4(9):733–738. doi:10.1002/dta.1402

This reference is revised as new data are published. Last reviewed 13 October 2026. Figures are taken from the cited sources or computed from the molecular formulas in the molarity reference. Nothing here is a protocol, and every measurement described is analytical work on a research material. Two instruments, two questions, and a certificate that answers both is the one worth keeping; the chromatogram and the mass are the two things we would ask any supplier for, ourselves included.